Next Generation Knee System
by Zimmer, Inc.
510(k) K933785
Next Generation Knee System is a Class II device made by Zimmer, Inc., cleared under FDA 510(k) K933785 on 30 Jan 1995, active as of 22 Oct 2026. 78 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SN | NumberK933785 | Date30 Jan 1995 | StatusActive |
Official records
- FDA 510(k) database K933785
Recalls
-
- Terminated Z-1038-2018
- Terminated Z-1040-2018
- Terminated Z-1039-2018
- Terminated Z-1042-2018
- Terminated Z-1041-2018
The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
-
- Terminated Z-2115-2017
Zimmer Biomet is conducting a lot-specific medical device field action for two lots of the NexGen Offset Stem Extension implants. Product complaints indicated that the size etched on the part and labeled on the product packaging were incorrect.
-
- Terminated Z-0835-2016
LDPE bag containing the implant adheres to the highly polished implant surface.
-
- Terminated Z-1159-2016
- Terminated Z-1151-2016
- Terminated Z-1147-2016
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
-
- Terminated Z-0590-2016
A product complaint was received stating that a carton labeled as a Precoat Stemmed Tibia Size 3 was opened for use and the contents were a non-augmented Stemmed Tibia Plate size 3.
-
- Terminated Z-0449-2016
Complaints were received from Thailand reporting that there was no white paper (Tyvek) seal on the inner package of P/N:00-5980-037-01 unit from lot #62460264
-
- Terminated Z-2552-2014
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
-
- Terminated Z-0480-2014
Two complaints were received in which it was reported that the stem would not lock into the tibia. Evaluation of the returned device found that the stemmed portion of the tibial plate would not hold the stem as the connection was too loose. The taper diameter was found to be oversized by .010.
-
- Terminated Z-1023-2014
Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.
-
- Terminated Z-1606-2012
- Terminated Z-1571-2012
- Terminated Z-1583-2012
- Terminated Z-1598-2012
- Terminated Z-1574-2012
- Terminated Z-1591-2012
- Terminated Z-1597-2012
- Terminated Z-1580-2012
- Terminated Z-1584-2012
- Terminated Z-1555-2012
- Terminated Z-1560-2012
- Terminated Z-1603-2012
- Terminated Z-1589-2012
- Terminated Z-1608-2012
- Terminated Z-1573-2012
- Terminated Z-1585-2012
- Terminated Z-1558-2012
- Terminated Z-1572-2012
- Terminated Z-1569-2012
- Terminated Z-1565-2012
- Terminated Z-1575-2012
- Terminated Z-1593-2012
- Terminated Z-1552-2012
- Terminated Z-1561-2012
- Terminated Z-1604-2012
- Terminated Z-1581-2012
- Terminated Z-1576-2012
- Terminated Z-1550-2012
- Terminated Z-1556-2012
- Terminated Z-1559-2012
- Terminated Z-1587-2012
- Terminated Z-1562-2012
- Terminated Z-1602-2012
- Terminated Z-1601-2012
- Terminated Z-1578-2012
- Terminated Z-1563-2012
- Terminated Z-1592-2012
- Terminated Z-1609-2012
- Terminated Z-1564-2012
- Terminated Z-1605-2012
- Terminated Z-1582-2012
- Terminated Z-1590-2012
- Terminated Z-1599-2012
- Terminated Z-1586-2012
- Terminated Z-1611-2012
- Terminated Z-1547-2012
- Terminated Z-1551-2012
- Terminated Z-1549-2012
- Terminated Z-1594-2012
- Terminated Z-1577-2012
- Terminated Z-1610-2012
- Terminated Z-1567-2012
- Terminated Z-1607-2012
- Terminated Z-1595-2012
- Terminated Z-1600-2012
- Terminated Z-1579-2012
- Terminated Z-1588-2012
- Terminated Z-1548-2012
- Terminated Z-1557-2012
- Terminated Z-1554-2012
- Terminated Z-1596-2012
- Terminated Z-1566-2012
- Terminated Z-1570-2012
- Terminated Z-1553-2012
- Terminated Z-1568-2012
Zimmer has received 5 complaints where a NexGen CR micro articular surface was used with a standard CR femur, even though the compatibility chart and product labeling indicates that these components are NOT compatible. Investigation identified additional instances where a CR micro articular surface and standard CR femur were sold together on the same purchase order. CR micro articular surfaces are
-
- Terminated Z-0609-2011
The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery. These complications may also require revision surgery.
-
- Terminated Z-0119-2011
- Terminated Z-0120-2011
Zimmer had determined that two lots of these tibial implants have incorrectly positioned or missing flange plugs that were not seated in the device upon receipt to the customer. A missing flange plug could lead to a delay of surgery or possibility of wear leading to more surgery.
Timeline
- Recall initiated FDA recalls Z-0835-2016
- Recall initiated FDA recalls Z-1159-2016
- Recall initiated FDA recalls Z-1151-2016
- Recall initiated FDA recalls Z-1147-2016
- Recall initiated FDA recalls Z-2115-2017
- Recall initiated FDA recalls Z-1038-2018
- Recall initiated FDA recalls Z-1040-2018
- Recall initiated FDA recalls Z-1039-2018
- Recall initiated FDA recalls Z-1042-2018
- Recall initiated FDA recalls Z-1041-2018
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JWH
- Device class
- Class II
- Regulation
- 21 CFR Part 888.3560
- Advisory committee
- Orthopedic
- FDA decision
- SN
Devices cleared under K933785
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| NexGen® 32-5988-152-15 | 32-5988-152-15 / 32-5988-152-15 | 00889024228627 | Length 152 mm | On the market |
| NexGen® 5966-04-12 | 00-5966-004-12 / 5966-04-12 | 00889024212336 | Height 12 mm | On the market |
| NexGen® 5970-17-02 | 00-5970-017-02 / 5970-17-02 | 00889024214194 | On the market | |
| NexGen® 5970-21-12 | 00-5970-021-12 / 5970-21-12 | 00889024214248 | Height 12 mm | On the market |
| NexGen® 5970-21-14 | 00-5970-021-14 / 5970-21-14 | 00889024214255 | Height 14 mm | On the market |
| NexGen® 5970-30-12 | 00-5970-030-12 / 5970-30-12 | 00889024214330 | Height 12 mm | On the market |
| NexGen® 5970-40-20 | 00-5970-040-20 / 5970-40-20 | 00889024214446 | Height 20 mm | On the market |
| NexGen® 5970-45-01 | 00-5970-045-01 / 5970-45-01 | 00889024214453 | On the market | |
| NexGen® 5970-55-02 | 00-5970-055-02 / 5970-55-02 | 00889024214545 | On the market | |
| NexGen® 5970-55-03 | 00-5970-055-03 / 5970-55-03 | 00889024214552 | On the market | |
| NexGen® 5972-65-38 | 00-5972-065-38 / 5972-65-38 | 00889024635401 | On the market | |
| NexGen® 5975-21-17 | 00-5975-021-17 / 5975-21-17 | 00889024215283 | Height 17 mm | No longer on the market |
| NexGen® 5975-40-17 | 00-5975-040-17 / 5975-40-17 | 00889024215405 | Height 17 mm | No longer on the market |
| NexGen® 5975-50-23 | 00-5975-050-23 / 5975-50-23 | 00889024215481 | Height 23 mm | No longer on the market |
| NexGen® 5976-30-12 | 00-5976-030-12 / 5976-30-12 | 00889024635487 | Height 12.00 mm | On the market |
| NexGen® 5976-30-14 | 00-5976-030-14 / 5976-30-14 | 00889024215573 | Height 14 mm | On the market |
| NexGen® 5976-31-10 | 00-5976-031-10 / 5976-31-10 | 00889024446847 | Height 10.00 mm | On the market |
| NexGen® 5976-40-12 | 00-5976-040-12 / 5976-40-12 | 00889024635531 | Height 12.00 mm | On the market |
| NexGen® 5976-40-17 | 00-5976-040-17 / 5976-40-17 | 00889024635555 | Height 17.00 mm | On the market |
| NexGen® 5976-40-23 | 00-5976-040-23 / 5976-40-23 | 00889024215665 | Height 23 mm | On the market |
| NexGen® 5976-41-17 | 00-5976-041-17 / 5976-41-17 | 00889024446939 | Height 17.00 mm | On the market |
| NexGen® 5976-50-10 | 00-5976-050-10 / 5976-50-10 | 00889024635579 | Height 10.00 mm | On the market |
| NexGen® 5976-50-12 | 00-5976-050-12 / 5976-50-12 | 00889024635586 | Height 12.00 mm | On the market |
| NexGen® 5980-27-01 | 00-5980-027-01 / 5980-27-01 | 00889024217553 | On the market | |
| NexGen® 5980-37-01 | 00-5980-037-01 / 5980-37-01 | 00889024217577 | On the market | |
| NexGen® 5980-57-04 | 00-5980-057-04 / 5980-57-04 | 00889024217652 | On the market | |
| NexGen® 5988-15-19 | 00-5988-015-19 / 5988-15-19 | 00889024221246 | Length 75 mm | On the market |
| NexGen® 5988-15-20 | 00-5988-015-20 / 5988-15-20 | 00889024221253 | Length 75 mm | On the market |
| NexGen® 5988-16-10 | 00-5988-016-10 / 5988-16-10 | 00889024221284 | Length 130 mm | On the market |
| NexGen® 5988-18-13 | 00-5988-018-13 / 5988-18-13 | 00889024221420 | Length 60 mm | On the market |
| NexGen® 5988-20-11 | 00-5988-020-11 / 5988-20-11 | 00889024221444 | Length 100 mm | On the market |
| NexGen® 5988-20-16 | 00-5988-020-16 / 5988-20-16 | 00889024221499 | Length 100 mm | On the market |
| NexGen® 5988-21-14 | 00-5988-021-14 / 5988-21-14 | 00889024221543 | Length 155 mm | On the market |
| NexGen® 90-5970-20-20 | 90-5970-020-20 / 90-5970-20-20 | 00889024429086 | Height 20.00 mm | On the market |
| NexGen® 90-5970-21-12 | 90-5970-021-12 / 90-5970-21-12 | 00889024644342 | Height 12.00 mm | On the market |
| NexGen® 90-5970-21-12 | 90-5970-021-12 / 90-5970-21-12 | 00889024263512 | Height 12 mm | On the market |
| NexGen® 90-5970-31-12 | 90-5970-031-12 / 90-5970-31-12 | 00889024644458 | Height 12.00 mm | On the market |
| NexGen® 90-5970-31-12 | 90-5970-031-12 / 90-5970-31-12 | 00889024429109 | Height 12.00 mm | On the market |
| NexGen® 90-5970-31-14 | 90-5970-031-14 / 90-5970-31-14 | 00889024644465 | Height 14.00 mm | On the market |
| NexGen® 90-5970-40-10 | 90-5970-040-10 / 90-5970-40-10 | 00889024263628 | Height 10 mm | On the market |
| NexGen® 90-5970-40-12 | 90-5970-040-12 / 90-5970-40-12 | 00889024644519 | Height 12.00 mm | On the market |
| NexGen® 90-5970-50-10 | 90-5970-050-10 / 90-5970-50-10 | 00889024263680 | Height 10 mm | On the market |
| NexGen® 90-5970-50-10 | 90-5970-050-10 / 90-5970-50-10 | 00889024644564 | Height 10.00 mm | On the market |
| NexGen® 90-5970-50-20 | 90-5970-050-20 / 90-5970-50-20 | 00889024644601 | Height 20.00 mm | On the market |
| NexGen® Legacy® 5992-15-02 | 00-5992-015-02 / 5992-15-02 | 00889024224612 | On the market | |
| NexGen® Legacy® 5996-03-10 | 00-5996-003-10 / 5996-03-10 | 00889024226326 | Height 10 mm | On the market |
| NexGen® Legacy® 5996-03-14 | 00-5996-003-14 / 5996-03-14 | 00889024226340 | Height 14 mm | On the market |
| NexGen® Legacy® 5996-08-17 | 00-5996-008-17 / 5996-08-17 | 00889024226555 | Height 17 mm | On the market |
| NexGen® Legacy® 5996-15-10 | 00-5996-015-10 / 5996-15-10 | 00889024226708 | Height 10 mm | On the market |
| NexGen® Legacy® 5996-18-14 | 00-5996-018-14 / 5996-18-14 | 00889024226944 | Height 14 mm | On the market |