# Nim 3.0

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/nim-3-0-gwf>
- Manufacturer: [Medtronic Xomed Inc](https://devicerts.com/manufacturers/medtronic-xomed-inc) (United States)
- Name in the registry: NIM 3.0
- Device class: Class II (US FDA)
- FDA product code: GWF

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K083124 · decision 2009-02-27 · decision SESE · status: Active

## Recalls

- Z-2343-2012 · Class II · terminated · initiated 2010-09-27: In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
- Z-2342-2012 · Class II · terminated · initiated 2010-09-27: In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
- Z-2344-2012 · Class II · terminated · initiated 2010-09-27: In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu

## Identifiers

- 510(k) Number: K083124 [FDA 510(k)]
- FDA product code: GWF [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.1870 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Nim 3.0, Medtronic Xomed Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/nim-3-0-gwf (data as of 2026-11-19).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-08
- Statuses computed as of: 2026-11-19
- JSON: <https://devicerts.com/devices/nim-3-0-gwf.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/nim-3-0-gwf
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.