NIM Neuro 8253402
NIM Neuro 8253402 is a nerve-locating system (FDA Class II) made by Medtronic Xomed Inc, cleared under FDA 510(k) K083124 on 27 Feb 2009 and listed in FDA GUDID since 23 Jan 2025 (UDI-DI 00763000835323), active as of 8 Nov 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00763000835323 | Date23 Jan 2025 | StatusListed |
| United States | FDA 510(k)SESE | NumberK083124 | Date27 Feb 2009 | StatusActive |
Official records
- AccessGUDID record 00763000835323
- FDA 510(k) database K083124
From the FDA record
- Device description
- MAINFRAME 8253402 NEURO 3.0 INTL EL ROHS
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K083124 decided 27 Feb 2009 Active
- FDA GUDID UDI-DI 00763000835323 listed 23 Jan 2025 Listed
Recalls
-
- Terminated Z-2343-2012
- Terminated Z-2342-2012
- Terminated Z-2344-2012
In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
Timeline
- Cleared FDA 510(k) K083124
- Recall initiated FDA recalls Z-2343-2012
- Recall initiated FDA recalls Z-2342-2012
- Recall initiated FDA recalls Z-2344-2012
- Listed FDA GUDID 00763000835323
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- NIM Neuro
- Generic name
- Nerve-locating system
- Model / reference
- 8253402
- Primary DI / UDI-DI
- 00763000835323
- 510(k) Number
- FDA product code
- GWF, ETN
- GMDN term
- Nerve-locating system
- Device class
- Class II
- Regulation number
- 882.1870, 874.1820
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code GWF (Stimulator, electrical, evoked response).
- SafeOp 3: Neural Informatix System by ALPHATEC SPINE, INC.
- NEUROPACK 8, MEE-4108 by NIHON KOHDEN AMERICA, INC.
- TES-1 CONSTANT VOLTAGE ELECTRICAL STIMULATOR by Cadwell Laboratories, Inc.
- ISIS 504281 by inomed Medizintechnik GmbH
- Digitimer DS7A by Digitimer, Ltd.
- 533624 by inomed Medizintechnik GmbH
- MEP Monitor by Tonica Elektronik A/S
- CANTATA EMG/EVOKED RESPONSE SYSTEM by Dantec Medical, Inc.
- EPOCHXP NEUROLOGICAL WORKSTATION, EPOCHXP LITE NEUROLOGICAL WORKSTATION by Axon Systems, Inc.
- CADWELL FOUR-/EIGHT-CHANNEL PREAMPLIFIER by Cadwell Laboratories, Inc.