Nim® 8253070
Nim 8253070 is a nerve-locating system (FDA Class II) made by Medtronic Xomed Inc, cleared under FDA 510(k) K083124 on 27 Feb 2009 and listed in FDA GUDID since 2 May 2016 (UDI-DI 00613994499578), active as of 8 Nov 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00613994499578 | Date2 May 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK083124 | Date27 Feb 2009 | StatusActive |
Official records
- AccessGUDID record 00613994499578
- FDA 510(k) database K083124
From the FDA record
- Device description
- FUSE KIT 8253070 NIM 3.0 2.5 AMP ROHS
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K083124 decided 27 Feb 2009 Active
- FDA GUDID UDI-DI 00613994499578 listed 2 May 2016 Listed
Recalls
-
- Terminated Z-2343-2012
- Terminated Z-2342-2012
- Terminated Z-2344-2012
In January 2010, Medtronic issued a recall of Medtronic NIM 3.0 Patient Interfaces after Medtronic identified a trend of open channel conditions (signal loss from the electrode). When the NIM 3.0 detects the signal loss from the electrode (i.e. open channel), it fails the electrode check and provides an audible and visual warning to the user which could also create a delay in the surgical procedu
Timeline
- Cleared FDA 510(k) K083124
- Recall initiated FDA recalls Z-2343-2012
- Recall initiated FDA recalls Z-2342-2012
- Recall initiated FDA recalls Z-2344-2012
- Listed FDA GUDID 00613994499578
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Nim®
- Generic name
- Nerve-locating system
- Model / reference
- 8253070
- Primary DI / UDI-DI
- 00613994499578
- 510(k) Number
- FDA product code
- ETN, GWF
- GMDN term
- Nerve-locating system
- Device class
- Class II
- Regulation number
- 874.1820, 882.1870
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code ETN (STIMULATOR, NERVE).
- 533666 by inomed Medizintechnik GmbH
- Invotec Single Channel Electrode 1 29-80500S by Invotec International, Inc.
- Nerve Monitor Electrodes/Stimulator Probes by Micromed Development Corp.
- 534672 by inomed Medizintechnik GmbH
- Invotec Single Channel Electrode 1 29-80500W by Invotec International, Inc.
- Neuro Check Device by Baxano, Inc.
- FREMAP Electrode DCNE-01-05X by Ad-Tech Medical Instrument Corp.
- XOMED-TREACE EMG ENDOTRACHEAL TUBEEN by Xomed-Treace, Inc.
- NEUROVISION INK PRINTED ENDOTRACHEAL TUBE ELECTRODE by Neurovision Medical Products Inc.
- CHECKPOINT by Ndi Medical, LLC