NOMAD Pro 2™ 0.850.0021
by Aribex, Inc.
NOMAD Pro 2 0.850.0021 (FP 0150) is a diagnostic x-ray system generator (FDA Class II; FDA product code EHD, Unit, X-Ray, Extraoral with Timer) made by Aribex, Inc., cleared under FDA 510(k) K081664 on 23 Jun 2008 and listed in FDA GUDID since 24 Sep 2016 (UDI-DI EKAV085000210), active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberEKAV085000210 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK081664 | Date23 Jun 2008 | StatusActive |
Official records
- FDA 510(k) database K081664
From the FDA record
- Device description
- Handheld X-ray
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K081664 decided 23 Jun 2008 Active
- FDA GUDID UDI-DI EKAV085000210 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-2716-2016
- Terminated Z-2717-2016
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
Timeline
- Cleared FDA 510(k) K081664
- Recall initiated FDA recalls Z-2716-2016
- Recall initiated FDA recalls Z-2717-2016
- Listed FDA GUDID EKAV085000210
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- NOMAD Pro 2™
- Generic name
- Diagnostic x-ray system generator, hand-held
- Model / Ref / Catalog
- 0.850.0021, FP 0150
- Primary DI / UDI-DI
- EKAV085000210
- 510(k) Number
- FDA product code
- EHD
- GMDN term
- Diagnostic x-ray system generator, hand-held
- Device class
- Class II
- Regulation number
- 872.1800
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code EHD (UNIT, X-RAY, EXTRAORAL WITH TIMER).
- Orthophos E 6707462 by SIRONA Dental Systems GmbH
- AZTECH 70 by VILLA SISTEMI MEDICALI S.P.A.
- MODEL 2001 CP, PHOTON I, DENTAL, INTRAORAL X-RAY by Belmont Equipment Corp.
- EXPLOR-X 70 by VILLA SISTEMI MEDICALI S.P.A.
- HELIODENT PLUS by SIRONA Dental Systems GmbH
- MINXRAY HF70D HIGH FREQUENCY PORTABLE DENTAL X-RAY UNIT by Minxray, Inc.
- COFFMAN TOMOGRAPH by Coffman Dental
- PORTABLE X-RAY SYSTEM by OSSTEM IMPLANT CO.,LTD
- XTG PRO 70 HYBRID Pro by DigiMed Co., Ltd.
- PM 2002 CC PROLINE by Planmeca Oy