Nordicbrainex Software Version 1.0
510(k) K092102
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK092102 | Date25 Aug 2009 | StatusActive |
Recalls
-
- Terminated Z-2045-2018
There is an error related to the relative geometry between fiber tracts in a fiber group and images. The error occurs in the following situations: A group of fibers has been selected using the VOI functionality and put into a fiber group. This is done with images series A visualized in the 3D viewer. Then a new images series B is visualized in the 3D viewer. If image series A and B have the same geometrical resolution (pixel sizes and slice distances), the coregistration matrix taking A to B will not be applied to the fiber group, and therefore the fiber group will in general not be positioned correctly on B in the 3D viewer. Furthermore, if the fiber group is exported as a new image series the fibers could be misplaced, and similarly if the fibers are visualized in the MPR on B, the positioning will not be correct. The misalignment will be equal to the rotations/translations necessary to align A with B.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- LLZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.2050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 23 Sep 2026
The device is a software product named Nordicbrainex Software, Version 1.0, developed by Nordicimaginglab AS. It is classified under the product code LLZ and has been determined to be substantially equivalent to a predicate device through a traditional 510(k) clearance process.
The software received FDA 510(k) clearance under K092102 on August 25, 2009, following submission on July 14, 2009. The review was conducted by the Radiology advisory committee (RA), and the device is manufactured at Mollendalsveien 65c, Bergen, Hordaland, Norway, with postal code 5009.
Other devices from Nordicimaginglab AS
Questions
AI-generated answers, 23 Sep 2026
What is Nordicbrainex Software Version 1.0 used for?
Nordicbrainex Software Version 1.0 is a software product designed for use in radiology to support medical image analysis, as indicated by its product code LLZ and review by the Radiology advisory committee during its FDA 510(k) clearance process.
How was Nordicbrainex Software Version 1.0 cleared for market release?
The software received FDA 510(k) clearance through a traditional submission process, demonstrating substantial equivalence to a predicate device, with clearance granted under K092102 on August 25, 2009, following review by the Radiology advisory committee.
Who manufactures Nordicbrainex Software Version 1.0 and where is it produced?
Nordicbrainex Software Version 1.0 is manufactured by Nordicimaginglab AS at their facility located at Mollendalsveien 65c, Bergen, Hordaland, Norway, with postal code 5009.