Noxboxi Nitric Oxide Delivery System
510(k) K171696
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK171696 | Date2 Oct 2018 | StatusActive |
Recalls
-
- Ongoing Z-3306-2026
Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
-
- Ongoing Z-3165-2026
Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
-
- Ongoing Z-0028-2026
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
-
- Ongoing Z-0029-2026
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
-
- Ongoing Z-2544-2025
Replacement of the internal Sample Pump due to high failure rates.
-
- Terminated Z-1667-2023
Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MRN
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5165
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 26 Sep 2026
The Noxboxi Nitric Oxide Delivery System is an apparatus designed for the delivery and monitoring of a constant concentration of nitric oxide. It is used to monitor nitric oxide, nitrogen dioxide, and oxygen levels within the inspiratory ventilator lines of patients undergoing nitric oxide therapy.
This device is intended for use by healthcare professionals and has received FDA 510(k) clearance. It is classified as a class 2 medical device under regulation number 868.5165.
Questions
AI-generated answers, 26 Sep 2026
What is the primary function of the NOxBOXi Nitric Oxide Delivery System?
This apparatus is designed to deliver a constant concentration of nitric oxide to patients undergoing therapy. It simultaneously monitors the levels of nitric oxide, nitrogen dioxide, and oxygen within the inspiratory ventilator lines to ensure accurate administration.
Who is authorized to operate the NOxBOXi system?
The device is intended for use exclusively by healthcare professionals. This ensures that the delivery and monitoring of nitric oxide therapy are managed by trained clinical staff capable of interpreting the system's readings and maintaining patient safety.
What is the regulatory status of this device?
The NOxBOXi Nitric Oxide Delivery System has received FDA 510(k) clearance. It is classified as a class 2 medical device under regulation number 868.5165, which pertains to apparatuses used for the delivery of nitric oxide.