## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K111576 · decision 2011-09-20 · decision SESE · status: Active

## Recalls

- Z-0209-2024 · Class II · ongoing · initiated 2023-09-28: Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
- Z-1549-2021 · Class II · terminated · initiated 2021-04-22: Potential capacitor failure which typically manifests as in inability to provide output on one or more channels. In some instances, users reported a charring/smoky odor or emission.
- Z-0664-2020 · Class II · terminated · initiated 2019-10-14: Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
- Z-1361-2016 · Class II · terminated · initiated 2013-10-09: NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.

## Identifiers

- 510(k) Number: K111576 [FDA 510(k)]
- FDA product code: GXD [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 882.4400 [FDA 510(k)]
- Advisory committee: Neurology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Nt 2000 Lesioning Generator, Neuro Therm, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/nt-2000-lesioning-generator-gxd (data as of 2026-11-22).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-22
- JSON: <https://devicerts.com/devices/nt-2000-lesioning-generator-gxd.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/nt-2000-lesioning-generator-gxd
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.