O-Arm Imaging System
by Medtronic Navigation, Inc. (Littleton)
510(k) K092564
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK092564 | Date17 Jun 2010 | StatusActive |
Recalls
-
- Terminated Z-1932-2019
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
-
- Terminated Z-0277-2019
- Terminated Z-0280-2019
- Terminated Z-0272-2019
- Terminated Z-0279-2019
- Terminated Z-0276-2019
- Terminated Z-0271-2019
- Terminated Z-0275-2019
- Terminated Z-0278-2019
- Terminated Z-0273-2019
- Terminated Z-0274-2019
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
-
- Terminated Z-0928-2018
Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report, Gantry motion, Network communication, System shutdown, and Early termination of 3D spin.
-
- Terminated Z-2082-2016
It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
-
- Terminated Z-0973-2014
Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
-
- Terminated Z-0926-2013
Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
-
- Terminated Z-2244-2012
A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.
-
- Terminated Z-0447-2011
Safety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapy
-
- Terminated Z-1849-2012
Software Version 3.1.1 does not meet navigational accuracy.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OXO
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent