O-Two E700, O-Two E600, O-Two E500
by O-Two Medical Technologies Inc
510(k) K141595
O-Two E700, O-Two E600, O-Two E500 is a medical device (FDA Class II; FDA product code BTL, Ventilator, Emergency, Powered (Resuscitator)) made by O-Two Medical Technologies Inc, cleared under FDA 510(k) K141595 on 16 Mar 2015, active as of 22 Oct 2026. 3 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK141595 | Date16 Mar 2015 | StatusActive |
Official records
- FDA 510(k) database K141595
From the FDA record
- Received by the FDA
- 13 Jun 2014
- FDA decision
- 16 Mar 2015 (Substantially Equivalent)
- FDA review time
- 276 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Terminated Z-0976-2019
- Terminated Z-0977-2019
- Terminated Z-0978-2019
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Timeline
- Cleared FDA 510(k) K141595
- Recall initiated FDA recalls Z-0976-2019
- Recall initiated FDA recalls Z-0977-2019
- Recall initiated FDA recalls Z-0978-2019
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- BTL
- FDA applicant
- O-Two Medical Technologies, Inc.
- Device class
- Class II
- Regulation
- 21 CFR Part 868.5925
- Advisory committee
- Anesthesiology
- FDA decision
- SESE Substantially Equivalent
Devices cleared under K141595
Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.
| UDI-DI | Size and differences | |||
|---|---|---|---|---|
| o_two e500 01EVE500 | 01EVE500 / 01EVE500 | 07540195033054 | Weight 2.4 kg; Height 155 mm; Length 200 mm; Length 250 mm | On the market |
| o_two e600 01EVE600 | 01EVE600 / 01EVE600 | 07540195033030 | Weight 2.4 kg; Height 155 mm; Width 200 mm; Length 250 mm | On the market |
| o_two e700 01EVE700 | 01EVE700 / 01EVE700 | 07540195033016 | Weight 2.4 kg; Height 155 mm; Width 200 mm; Length 250 mm | On the market |
Similar devices
Other devices filed under FDA product code BTL (Ventilator, Emergency, Powered (Resuscitator)).
- AUTOVENT MODEL 4000 by Allied Healthcare Products, Inc.
- BABYPAC by Smiths Medical International, Ltd.
- RESUSCITATOR MODELS 10400/10500/10600/10700/10800 by U.O. Equipment Co.
- OHMEDA - LOGIC 07A TRANSPORT VENTILATOR by Ohmeda Medical
- AUTOVENT 4000 WITH CPAP by Allied Healthcare Products, Inc.
- Resusci Flow w/Blender/Rail mounting type 60363 by Atom Medical Corporation
- VENTRAIN by Dolphys Medical B.V.
- MOVES SLC by Thornhill Research Inc
- LIFEBORNE INFANT RESUSCITATOR by INTERNATIONAL BIOMEDICAL, LTD.
- MEDUtrigger 20900 by WEINMANN Emergency Medical Technology GmbH