  "format": "devicerts.device.v1",
  "cite_as": "O-Two E700, O-Two E600, O-Two E500, O-Two Medical Technologies Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl (data as of 2026-11-27).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl",
    "json_url": "https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl.json",
    "markdown_url": "https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-27",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "O-Two E700, O-Two E600, O-Two E500",
    "registry_name": "O-TWO E700, O-TWO E600, O-TWO E500",
    "url": "https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl",
    "manufacturer": {
      "name": "O-Two Medical Technologies Inc",
      "url": "https://devicerts.com/manufacturers/o-two-medical-technologies-inc",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "Canada"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K141595",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "BTL"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K141595",
      "number_label": null,
      "device_udi_di": null,
      "date": "2015-03-16",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0976-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-02-12",
      "reason": "There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.",
      "applies": "this device",
      "via": "K141595"
    },
    {
      "number": "Z-0977-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-02-12",
      "reason": "There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.",
      "applies": "this device",
      "via": "K141595"
    },
    {
      "number": "Z-0978-2019",
      "classification": "Class I",
      "status": "terminated",
      "initiated": "2019-02-12",
      "reason": "There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.",
      "applies": "this device",
      "via": "K141595"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K141595",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "BTL",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 868.5925",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Anesthesiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}