# O-Two E700, O-Two E600, O-Two E500

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl>
- Manufacturer: [O-Two Medical Technologies Inc](https://devicerts.com/manufacturers/o-two-medical-technologies-inc) (FDA registration: Canada)
- Name in the registry: O-TWO E700, O-TWO E600, O-TWO E500
- Device class: Class II (US FDA)
- FDA product code: BTL

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K141595 · decision 2015-03-16 · decision SESE · status: Active

## Recalls

- Z-0976-2019 · Class I · terminated · initiated 2019-02-12: There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
- Z-0977-2019 · Class I · terminated · initiated 2019-02-12: There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
- Z-0978-2019 · Class I · terminated · initiated 2019-02-12: There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

## Identifiers

- 510(k) Number: K141595 [FDA 510(k)]
- FDA product code: BTL [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 868.5925 [FDA 510(k)]
- Advisory committee: Anesthesiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: O-Two E700, O-Two E600, O-Two E500, O-Two Medical Technologies Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl (data as of 2026-11-27).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-27
- JSON: <https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/o-two-e700-o-two-e600-o-two-e500-btl
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.