O640 Upgrade H3100ZC
by GE Medical Systems Israel, Functional Imaging
O640 Upgrade H3100ZC is a full-body CT system (FDA Class II) made by GE Medical Systems Israel, Functional Imaging, cleared under FDA 510(k) K121019 on 19 Apr 2012 and listed in FDA GUDID since 28 Jun 2017 (UDI-DI 00840682123730), active as of 23 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00840682123730 | Date28 Jun 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK121019 | Date19 Apr 2012 | StatusActive |
Official records
- AccessGUDID record 00840682123730
- FDA 510(k) database K121019
From the FDA record
- Device description
- D630 TO O640 UPGRDAE COLLECTOR
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K121019 decided 19 Apr 2012 Active
- FDA GUDID UDI-DI 00840682123730 listed 28 Jun 2017 Listed
Recalls
-
- Ongoing Z-0896-2025
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
-
- Ongoing Z-1056-2023
A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.
-
- Terminated Z-0858-2021
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
-
- Terminated Z-1266-2019
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
-
- Terminated Z-0367-2014
GE Healthcare has recently become aware of a potential safety issue related to a SPECT scan procedure used for pulmonary imaging. Contact between the Detector (GE Healthcare Discovery NM/CT 670, and/or Optima NM/CT 640, and/or Discovery NM630, and/or Brivo NM 615 all configurations) and the Patient's elbow may occur during SPECT protocols using 450 angular steps in scenarios when a Patients a
-
- Terminated Z-1793-2013
GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. UPDATE: On July 03, 2013 GE notified hospitals that they are recalling several Nuclear Medicine Imaging Systems because serious injuries or deaths could occur due to the failure mode associated with this re
Timeline
- Cleared FDA 510(k) K121019
- Recall initiated FDA recalls Z-1793-2013
- Recall initiated FDA recalls Z-0367-2014
- Listed FDA GUDID 00840682123730
- Recall initiated FDA recalls Z-1266-2019
- Recall initiated FDA recalls Z-0858-2021
- Recall initiated FDA recalls Z-1056-2023
- Recall initiated FDA recalls Z-0896-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- O640 Upgrade
- Generic name
- Full-body CT system
- Model / reference
- H3100ZC
- Primary DI / UDI-DI
- 00840682123730
- 510(k) Number
- FDA product code
- JAK, KPS
- GMDN term
- Full-body CT system
- Device class
- Class II
- Regulation number
- 892.1750, 892.1200
- GUDID distribution status
- In Commercial Distribution
Other devices from GE Medical Systems Israel, Functional Imaging
See all devices from GE Medical Systems Israel, Functional Imaging →