# Ocean Chest Drain

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/ocean-chest-drain-kdq>
- Manufacturer: [Atrium Medical Corporation](https://devicerts.com/manufacturers/atrium-medical-corporation) (United States)
- Name in the registry: OCEAN CHEST DRAIN
- Device class: Class II (US FDA)
- FDA product code: KDQ

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K043582 · decision 2005-01-21 · decision SESE · status: Active

## Recalls

- Z-0486-2024 · Class II · ongoing · initiated 2023-09-05 · matched through shared submission K043582, may concern other models: Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
- Z-0235-2023 · Class II · ongoing · initiated 2022-10-18: Warehouse experienced temperature excursions in July and August 2021. It was determined that there were no associated harms to patients.
- Z-1607-2022 · Class II · ongoing · initiated 2022-05-25: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- Z-0307-2022 · Class II · ongoing · initiated 2021-10-22: Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient
- Z-1210-2017 · Class II · terminated · initiated 2016-12-23: Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
- Z-1209-2017 · Class II · terminated · initiated 2016-12-23: Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
- Z-1206-2017 · Class II · terminated · initiated 2016-12-23: Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
- Z-1205-2017 · Class II · terminated · initiated 2016-12-23: Chest drains manual vent valve located at the top of affected chest drains could potentially not re-seal after being depressed.
- Z-1938-2015 · Class II · terminated · initiated 2015-05-07 · matched through shared submission K043582, may concern other models: Chest drains have a missing o-ring
- Z-1935-2015 · Class II · terminated · initiated 2015-05-07 · matched through shared submission K043582, may concern other models: Chest drains have a missing o-ring
- Z-1937-2015 · Class II · terminated · initiated 2015-05-07 · matched through shared submission K043582, may concern other models: Chest drains have a missing o-ring
- Z-1936-2015 · Class II · terminated · initiated 2015-05-07 · matched through shared submission K043582, may concern other models: Chest drains have a missing o-ring
- Z-1959-2014 · Class II · terminated · initiated 2014-04-28 · matched through shared submission K043582, may concern other models: Product that was previously recalled because the Chest Drain tubing of the ATS Blood recovery may leak or disconnect, was released for distribution in error.
- Z-0743-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0740-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0738-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0736-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0739-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0735-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0742-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0741-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
- Z-0737-2014 · Class II · terminated · initiated 2013-11-19 · matched through shared submission K043582, may concern other models: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

## Identifiers

- 510(k) Number: K043582 [FDA 510(k)]
- FDA product code: KDQ [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.6740 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Ocean Chest Drain, Atrium Medical Corporation. deviCERTS medical device catalogue. https://devicerts.com/devices/ocean-chest-drain-kdq (data as of 2026-11-21).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-16
- Statuses computed as of: 2026-11-21
- JSON: <https://devicerts.com/devices/ocean-chest-drain-kdq.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.