OEC 9900 ELITE
by Ge Healthcare Surgery GE Oec Medical Systems, Inc. 2 variants
510(k) K122234
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK122234 | Date16 Aug 2012 | StatusActive |
Recalls
-
- Ongoing Z-2342-2025
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
-
- Ongoing Z-0705-2025
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
-
- Ongoing Z-2277-2023
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
-
- Ongoing Z-2872-2020
GEHC Surgery became aware that some users of the 9800 and 9900 fluoroscopic mobile C-Arms (referred to as the system here on) within the United States had requested that a Field Service Engineer (FSE) change the dose value on the live update screen from AKR / CAK to DAP. All systems sold to customers in the US that leave GEHC Surgery (the place of manufacture) are configured to display AKR / CAK when they leave the factory. An FSE changing the dose display, per the customer request, to display DAP instead does not comply to 21CFR 1020.32(k), which requires fluoroscopic equipment manufactured on or after June 10, 2006 to display continuous updates of AKR and CAK values.
-
- Terminated Z-1060-2017
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
-
- Terminated Z-0975-2017
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 22 Sep 2026
The OEC 9900 ELITE is an Interventional Fluoroscopic X-Ray System. It is a mobile C-arm device used for surgical imaging and interventional procedures.
The device is supplied as a complete system. It is intended for use in clinical, surgical, and vascular settings. The device is not reusable.
Other devices from Ge Healthcare Surgery GE Oec Medical Systems, Inc.
Questions
AI-generated answers, 22 Sep 2026
What is the OEC 9900 ELITE used for?
The OEC 9900 ELITE is a mobile C-arm device designed for surgical imaging and interventional procedures. It is intended for use in clinical, surgical, and vascular settings to provide real-time X-ray imaging during procedures.
How is the OEC 9900 ELITE supplied?
The device is supplied as a complete system, including all necessary components for interventional fluoroscopic imaging. It is not reusable and is intended for single-use procedures in clinical environments.
What is the regulatory status of the OEC 9900 ELITE?
The OEC 9900 ELITE has received FDA 510(k) clearance with K number K122234. It is classified as a Class 2 device under the product code OWB and regulation number 892.1650, deemed substantially equivalent to other devices on the market.
What are the specifications of the OEC 9900 ELITE?
The OEC 9900 ELITE is an Interventional Fluoroscopic X-Ray System with mobile C-arm functionality. It was manufactured by GE Healthcare Surgery GE Oec Medical Systems, Inc. and received FDA clearance on August 16, 2012.