Ohmeda Medical Omnibed
by Ohmeda Medical 510(k)
Active
510(k) K993407
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK993407 | Date5 Jan 2000 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-2236-2020
Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
-
- Terminated Z-0041-2020
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
-
- Terminated Z-1509-2017
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
-
- Terminated Z-1047-05
Neonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.
Specification
- 510(k) Number
- FDA Product Code
- FMZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 880.5400
- Advisory Committee
- General Hospital
- FDA Decision
- SESE Substantially Equivalent