Olympus AU5400 Clinical Chemistry Analyzer
by Olympus America, Inc. 510(k)
Active
510(k) K011720
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK011720 | Date25 Jun 2001 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-0152-2014
The recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE REF solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.
-
- Terminated Z-1430-2013
Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.
-
- Terminated Z-0974-2012
- Terminated Z-0977-2012
The recall was initiated because Beckman has confirmed that the AU Clinical Chemistry Analyzers may experience cuvette overflow.
Specification
- 510(k) Number
- FDA Product Code
- CIX
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1035
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent