Skip to content

Olympus Hyf-P Flexible Hysterofiberscope

by Olympus Corp.

510(k) K891451

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK891451 Date2 Oct 1989ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-0335-2024 FDA Class II initiated 25 Sep 2023

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
HIH
Device Class
Class II
Regulation
21 CFR Part 884.1690
Advisory Committee
Obstetrics/Gynecology
FDA Decision
SESE Substantially Equivalent

Other devices from Olympus Corp.

See all devices from Olympus Corp. →