Olympus Hyf-P Flexible Hysterofiberscope
510(k) K891451
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK891451 | Date2 Oct 1989 | StatusActive |
Recalls
-
- Ongoing Z-0335-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HIH
- Device Class
- Class II
- Regulation
- 21 CFR Part 884.1690
- Advisory Committee
- Obstetrics/Gynecology
- FDA Decision
- SESE Substantially Equivalent