Olympus LDL Cholesterol Reagent OSR6196/OSR6296; Olympus LDL Cholesterol Calibrator ODC0024

by Olympus America, Inc.

510(k) K040471

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

Recall on shared 510(k) K040471; other models may be affected

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK040471 Date10 Mar 2004ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-1135-2026 FDA Class II initiated 8 Dec 2025 on shared 510(k) K040471

    Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JIT
Device Class
Class II
Regulation
21 CFR Part 862.1150
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

This device is a secondary calibrator and reagent set for the quantitative determination of low-density lipoprotein (LDL) cholesterol in human serum and plasma. The reagents enable accurate measurement of LDL cholesterol levels, supporting clinical assessments of cardiovascular risk. The calibrator ensures traceability and consistency in analytical results across laboratory testing systems.

The LDL cholesterol reagent and calibrator are intended for use in clinical laboratory settings by trained personnel. Supplied as single-use reagents, they require no sterilisation beyond manufacturer-provided packaging. The device complies with FDA 510(k) clearance (K040471) under Special Controls (product code JIT) and is classified as Class II. No MRI safety or reusable components are specified.

Other devices from Olympus America, Inc.

See all devices from Olympus America, Inc. →

Questions

AI-generated answers, 1 Oct 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Olympus LDL cholesterol reagent and calibrator used for?

    The Olympus LDL cholesterol reagent (OSR6196/OSR6296) and calibrator (ODC0024) are designed for the quantitative determination of low-density lipoprotein (LDL) cholesterol in human serum and plasma. This measurement supports clinical assessments of cardiovascular risk by providing accurate LDL levels, which are critical for evaluating patient risk of heart disease.

  • How are the LDL cholesterol reagent and calibrator supplied, and do they require sterilization?

    The LDL cholesterol reagent and calibrator are supplied as single-use items, meaning each kit is intended for one-time use only. No additional sterilization is required beyond the manufacturer-provided packaging, which ensures sterility for clinical laboratory use.

  • Are the Olympus LDL cholesterol reagent and calibrator reusable, or are they intended for single-use?

    These reagents and calibrator are explicitly designed for single-use only. This ensures consistency, accuracy, and safety in laboratory testing, as reuse could compromise sterility and result integrity.

  • What regulatory clearance does the Olympus LDL cholesterol reagent and calibrator hold?

    The Olympus LDL cholesterol reagent and calibrator hold FDA 510(k) clearance under Special Controls (product code JIT) and are classified as Class II devices. This clearance confirms that the device meets FDA requirements for safety and effectiveness as a secondary calibrator and reagent set for LDL cholesterol measurement.

  • What is the purpose of the LDL cholesterol calibrator in this set?

    The LDL cholesterol calibrator (ODC0024) ensures traceability and consistency in analytical results across laboratory testing systems. It provides a standardized reference point for the reagent (OSR6196/OSR6296), allowing laboratories to maintain accuracy and reliability in LDL cholesterol measurements over time.