Olympus Lf-Tp and Lf-Dp Tracheal Intubation Fiberscopes, Accessories and Ancillary Equipment
510(k) K981543
Active
510(k)
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK981543 | Date8 Jun 1998 | StatusActive |
Recalls
-
- Ongoing Z-0557-2024
Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification.
-
- Ongoing Z-0338-2024
- Ongoing Z-0344-2024
- Ongoing Z-0332-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
-
- Ongoing Z-2504-2023
Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- EOQ
- Device Class
- Class II
- Regulation
- 21 CFR Part 874.4680
- Advisory Committee
- Ear, Nose, Throat
- FDA Decision
- SESE Substantially Equivalent