Olympus Lyophilized Chemistry Calibrators DR0070 and DR0071
510(k) K043460
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK043460 | Date31 Jan 2005 | StatusActive |
Recalls
-
- Terminated Z-0605-2018
Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JIX
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1150
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent