Olympus Tjf-140R/Jf-140R Evis Duodenovideoscopes and Accessories, Olympus Maj-311/Maj-411 Distal Covers
510(k) K980465
Active
510(k)
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK980465 | Date21 Apr 1998 | StatusActive |
Recalls
-
- Ongoing Z-0341-2024
- Ongoing Z-0342-2024
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- FDT
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.1500
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent