Olympus Unsaturated Iron Binding Capcity Reagent, OSR6124/0SR6524
510(k) K041793
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK041793 | Date16 Jul 2004 | StatusActive |
Recalls
-
- Ongoing Z-2264-2025
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
-
- Terminated Z-3028-2020
The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JMO
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1415
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent