Skip to content

Olympus Unsaturated Iron Binding Capcity Reagent, OSR6124/0SR6524

by Olympus America, Inc.

510(k) K041793

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK041793 Date16 Jul 2004ClearedStatusActive

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-2264-2025 FDA Class II initiated 26 Jun 2025

    Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

    • Terminated Z-3028-2020 FDA Class II initiated 18 Aug 2020

    The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JMO
Device Class
Class I
Regulation
21 CFR Part 862.1415
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Olympus America, Inc.