Oncentra, Oncentra External Beam, Oncentra Brachy
by Elekta Inc. D/B/A Nucletron Corporation
510(k) K121448
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK121448 | Date2 Jul 2012 | StatusActive |
Recalls
-
- Terminated Z-0987-2017
Cross profile for Varian 60 degree wedge shows "horns."
-
- Terminated Z-0604-2016
When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
-
- Terminated Z-1712-2015
Incorrect dose calculation for Regions of Interest (ROIs) defined on a secondary image series.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MUJ
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent