One Step Pregnancy Test 10001
One Step Pregnancy Test 10001 (100-01) is a total human chorionic gonadotropin IVD (FDA Class II; FDA product code LCX, Kit, Test, Pregnancy, Hcg, Over The Counter) made by Universal Meditech Inc., listed in FDA GUDID since 15 Oct 2021 (UDI-DI B512100011), registry data as of 23 Oct 2026. One FDA recall is linked to it.
Listed
Valid until
Not applicable
FDA recalls
1 ongoingDocuments
None
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberB512100011 | Date15 Oct 2021 | StatusListed |
From the FDA record
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Recalls
-
- Ongoing Z-0042-2024
Universal Meditech Inc. was violatively distributing PrestiBio" Breastmilk Alcohol Test Strip, DiagnosUS SARS-CoV-2 Antibody (IgG/IgM) Test, HealthyWiser KetoFast" Ketone Test Strips and PrestiBio" Ketone Test Strips without marketing authorization. UMI is also recall their tests because they are going out of business and would not be able to continue fulfilling any post-market responsibilities of these distributed products. UMI claims to hold 510(k)s for the other devices, purchased the intellectual property in 2015 from the previous 510(k) holder, but FDA has been unable to verify this information. The devices may have unknown performance characteristics because the firm is unable to provide any documentation that the devices were manufactured in conformance with the Quality System regulation, including documentation (e.g., DHF) that the distributed devices had not been modified since original clearance in a way that could impact their safety and effectiveness, documentation of controlled storage temperature/humidity, and post-market surveillance documentation, such as complaints and adverse events.
Timeline
- Listed FDA GUDID B512100011
- Recall initiated FDA recalls Z-0042-2024
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — GUDID listing Listed
Specification
- Trade name
- One Step Pregnancy Test
- Generic name
- Total human chorionic gonadotropin IVD, kit, rapid ICT, clinical
- Model / Ref / Catalog
- 10001, 100-01
- Primary DI / UDI-DI
- B512100011
- FDA product code
- LCX
- GMDN term
- Total human chorionic gonadotropin IVD, kit, rapid ICT, clinical
- Device class
- Class II
- Regulation number
- 862.1155
- GUDID distribution status
- In Commercial Distribution
Similar devices in this category
Other devices filed under Total human chorionic gonadotropin IVD, kit, rapid ICT, clinical.