Opti Lion Electrolyte Analyzer, Model GD7200
510(k) K052027
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK052027 | Date29 Sep 2005 | StatusActive |
Recalls
-
- Terminated Z-0029-2014
OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.
-
- Terminated Z-2558-2011
The values reported by the chloride sensors may report erroneous values below specification on aqueous controls and potentially with patient samples of whole blood and serum/plasma.
-
- Terminated Z-1083-2010
Outer box labeling shows an incorrect storage temperature range.
-
- Terminated Z-0119-2008
Incorrect results/stability failure; Actual device stability studies show sodium and chloride sensors age differently than anticipated when stored at the high end of labeled storage, therefore subsequent measurements give results outside the allowed error allowance before the 6 month expiration date shown on the device.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CGZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 862.1170
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent