Skip to content

Optical Guidance Platform

by Varian Medical Systems, Inc.

510(k) K071360

Optical Guidance Platform is a medical device (FDA Class II; FDA product code IYE, Accelerator, Linear, Medical) made by Varian Medical Systems, Inc., cleared under FDA 510(k) K071360 on 8 Jun 2007, active as of 22 Oct 2026. 5 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

5 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK071360 Date8 Jun 2007ClearedStatusActive

Official records

From the FDA record

Received by the FDA
15 May 2007
FDA decision
8 Jun 2007 (Substantially Equivalent)
FDA review time
24 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K071360 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

5 recalls · FDA

    • Terminated Z-0332-2012 FDA Unclassified (pre-2012) initiated 4 Oct 2011

    An anomaly has been identified with the Head Ring posts used by both the Optical Guidance Platform FrameArray module and the Floorstand where the head ring posts may be damaged due to excessive mechanical stress resulting in possible failure during usage.

    • Terminated Z-2200-2011 FDA Unclassified (pre-2012) initiated 30 Mar 2011

    A software anomaly has been identified with the Optical Guidance Platform (OGP) Software v2.6 and v2.6.1 where the transfer of datasets from treatment planning systems other than the FastPlan system result in a lateral offset error.

    • Terminated Z-2063-2011 FDA Unclassified (pre-2012) initiated 3 Mar 2011

    The optical guidance platform may not be properly enforcing a 24 hour time limit between optical camera recalibrations.

    • Terminated Z-2176-2011 FDA Unclassified (pre-2012) initiated 3 Mar 2011

    The anomaly that has been identified with the Optical Guidance Platform (OGP) software may not be always displaying the correct transfer date on the patient file.

    • Terminated Z-0888-2011 FDA Unclassified (pre-2012) initiated 26 Jan 2009

    Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact.

Timeline

6 dated events

  1. Cleared FDA 510(k) K071360
  2. Recall initiated FDA recalls Z-0888-2011
  3. Recall initiated FDA recalls Z-2063-2011
  4. Recall initiated FDA recalls Z-2176-2011
  5. Recall initiated FDA recalls Z-2200-2011
  6. Recall initiated FDA recalls Z-0332-2012

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
IYE
Device class
Class II
Regulation
21 CFR Part 892.5050
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).

Other devices from Varian Medical Systems, Inc.

See all devices from Varian Medical Systems, Inc. →