Optima Coil System OPTI0306CST10
by Balt USA LLC
Optima Coil System OPTI0306CST10 is a neurovascular embolization coil (FDA Class II) made by Balt USA LLC, cleared under FDA 510(k) K200030 on 1 Feb 2020 and listed in FDA GUDID since 27 Feb 2020 (UDI-DI 00818053024825), active as of 23 Oct 2026. 5 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
5 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00818053024825 | Date27 Feb 2020 | StatusListed |
| United States | FDA 510(k)SESE | NumberK200030 | Date1 Feb 2020 | StatusActive |
Official records
- AccessGUDID record 00818053024825
- FDA 510(k) database K200030
From the FDA record
- Device description
- The Optima Coil System is indicated for endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations (AVMs) and arteriovenous fistulae. The Optima Coil System is also indicated for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolization in the peripheral vasculature. The Optima Coil System comprises of implant coil and a pusher assembly. The Optima Coil System is designed fr use with the XCEL Detachment Controller. The Optima coil is detached thermally with the XCEL Detachment Controller.
- MRI safety
- MR Conditional
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied sterile
Across registries
- FDA 510(k) K200030 decided 1 Feb 2020 Active
- FDA GUDID UDI-DI 00818053024825 listed 27 Feb 2020 Listed
Recalls
-
- Ongoing Z-1988-2025
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
-
- Ongoing Z-1041-2024
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
-
- Ongoing Z-0103-2023
Product pouch label does not match up with carton label
-
- Ongoing Z-1680-2022
Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.
-
- Ongoing Z-0660-2022
Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.
Timeline
- Cleared FDA 510(k) K200030
- Listed FDA GUDID 00818053024825
- Recall initiated FDA recalls Z-0660-2022
- Recall initiated FDA recalls Z-1680-2022
- Recall initiated FDA recalls Z-0103-2023
- Recall initiated FDA recalls Z-1041-2024
- Recall initiated FDA recalls Z-1988-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Optima Coil System
- Generic name
- Neurovascular embolization coil
- Model / Ref / Catalog
- OPTI0306CST10
- Primary DI / UDI-DI
- 00818053024825
- 510(k) Number
- FDA product code
- KRD, HCG
- GMDN term
- Neurovascular embolization coil
- Device class
- Class II
- Regulation number
- 870.3300, 882.5950
- GUDID distribution status
- In Commercial Distribution
- MRI safety
- MR Conditional
Similar devices
Other devices filed under FDA product code KRD (Device, Vascular, For Promoting Embolization).
- GEL-BEAD EMBOLIZATION SPHERES by Vascular Solutions, Inc.
- ENDOLUMINAL OCCLUSION SYSTEM (EOS) by Artventive Medical Group, Inc.
- FIBERED IDC OCCLUSION SYSTEM by Boston Scientific
- COIL POSITIONER SET by Cook, Inc.
- Ruby XL System by PENUMBRA, INC.
- Interlock™-35 M001363820 by Boston Scientific
- IMPEDE Embolization Plug, 5mm, IMPEDE Embolization Plug, 7mm, IMPEDE Embolization Plug, 10mm by Shape Memory Medical Inc
- Solus Gold Embolization Device by Artio Medical, Inc.
- COOK MREYE EMBOLIZATION COILS by Cook, Inc.
- AZUR PERIPHERAL HYDROCOIL ENDOVASCULAR EMBOLIZATION SYSTEM by MICROVENTION INC.