Optima Coil System
by Blockade Medical, LLC (D.B.A. Balt Usa)
510(k) K172390
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK172390 | Date18 Feb 2018 | StatusActive |
Recalls
-
- Ongoing Z-1988-2025
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
-
- Ongoing Z-1041-2024
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
-
- Ongoing Z-0103-2023
Product pouch label does not match up with carton label
-
- Ongoing Z-1680-2022
Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HCG
- Device Class
- Class II
- Regulation
- 21 CFR Part 882.5950
- Advisory Committee
- Neurology
- FDA Decision
- SESE Substantially Equivalent