# Optima Coil System OPTI0101CSS10

> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/optima-coil-system-opti0101css10>
- Manufacturer: [Balt USA LLC](https://devicerts.com/manufacturers/balt-usa-llc) (United States)
- Device class: Class III (EU MDD); Class II (US FDA)
- FDA product code: HCG
- EMDN: Z12109001 DETACHABLE COIL EQUIPMENT FOR BRAIN ANEURYSMS
- GMDN: Neurovascular embolization coil (60940)

## Registrations and authorizations

- **EUDAMED** · CE/MDD · European Union · Basic UDI-DI B-00818053025709 · UDI-DI 00818053025709 · registered 2026-10-13 · status: Active
- **GUDID** · GUDID listing · United States · UDI-DI 00818053025709 · listed 2020-02-26 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K172390 · decision 2018-02-18 · decision SESE · status: Active · from GUDID

## Recalls

- Z-1988-2025 · Class II · ongoing · initiated 2025-05-08: Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
- Z-1041-2024 · Class II · ongoing · initiated 2024-01-03: Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
- Z-0103-2023 · Class II · ongoing · initiated 2022-09-07: Product pouch label does not match up with carton label
- Z-1680-2022 · Class II · ongoing · initiated 2022-07-26: Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.

## Identity

- Trade name: Optima Coil System [FDA GUDID (AccessGUDID)]
- Generic name: Neurovascular embolization coil [FDA GUDID (AccessGUDID)]
- Device name: Neurovascular embolization device [EUDAMED]
- Model / reference: OPTI0101CSS10 [EUDAMED, FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00818053025709 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: B-00818053025709 [EUDAMED]
- 510(k) Number: K172390 [FDA GUDID (AccessGUDID)]
- FDA product code: HCG [FDA GUDID (AccessGUDID)]
- EMDN code: Z12109001 [EUDAMED]
- GMDN term: Neurovascular embolization coil (60940) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class III [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 882.5950 [FDA GUDID (AccessGUDID)]
- EU registration type: Legacy device (under a directive) [EUDAMED]
- EU market of registration: France [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- MRI safety: MR Conditional [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Implantable: Yes [EUDAMED, FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Measuring function: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Optima Coil System OPTI0101CSS10, Balt USA LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/optima-coil-system-opti0101css10 (data as of 2026-11-17).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-24
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/optima-coil-system-opti0101css10.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/optima-coil-system-opti0101css10
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.