  "format": "devicerts.device.v1",
  "cite_as": "Optima CT660, GE Medical Systems, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/optima-ct660-jak-k131576 (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/optima-ct660-jak-k131576",
    "json_url": "https://devicerts.com/devices/optima-ct660-jak-k131576.json",
    "markdown_url": "https://devicerts.com/devices/optima-ct660-jak-k131576.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/optima-ct660-jak-k131576",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Optima CT660",
    "registry_name": "OPTIMA CT660",
    "url": "https://devicerts.com/devices/optima-ct660-jak-k131576",
    "manufacturer": {
      "name": "GE Medical Systems, LLC",
      "url": "https://devicerts.com/manufacturers/ge-medical-systems-llc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K131576",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K131576",
      "number_label": null,
      "device_udi_di": null,
      "date": "2013-08-30",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0608-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-11-05",
      "reason": "GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-1900-2021",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-06-14",
      "reason": "Improperly loaded software options may result in additional X -ray radiation exposure to the patient.",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2583-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-24",
      "reason": "Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible  Power Supply).",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2584-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-05-24",
      "reason": "Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible  Power Supply).",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2873-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-04-27",
      "reason": "GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to\r\nService Engineers working inside the gantry.",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2870-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-04-27",
      "reason": "GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2181-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-05-19",
      "reason": "It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.",
      "applies": "this device",
      "via": "K131576"
    },
    {
      "number": "Z-2182-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2017-05-19",
      "reason": "It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.",
      "applies": "this device",
      "via": "K131576"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K131576",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}