Optima MR 430S
510(k) K123068
Optima MR 430S is a medical device (FDA Class II; FDA product code LNH, System, Nuclear Magnetic Resonance Imaging) made by GE Medical Systems, LLC, cleared under FDA 510(k) K123068 on 13 Nov 2012, active as of 22 Oct 2026. 2 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK123068 | Date13 Nov 2012 | StatusActive |
Official records
- FDA 510(k) database K123068
From the FDA record
- Received by the FDA
- 1 Oct 2012
- FDA decision
- 13 Nov 2012 (Substantially Equivalent)
- FDA review time
- 43 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-0620-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Terminated Z-1928-2016
The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.
Timeline
- Cleared FDA 510(k) K123068
- Recall initiated FDA recalls Z-1928-2016
- Recall initiated FDA recalls Z-0620-2022
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- LNH
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1000
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code LNH (System, Nuclear Magnetic Resonance Imaging).
- M2 WRIST MRI SYSTEM by Aspect Imaging, Ltd.
- KINEMATIC MRI MOVIEMAKER by Resonex, Inc.
- Teslacon NMR Imaging System by Technicare Corp.
- VISTA HPQ SYSTEM by Philips Medical Systems (Cleveland), Inc.
- Vantage Fortian/Orian 1.5T, MRT-1550, V9.0 with AiCE Reconstruction Processing Unit for MR by Canon Medical Systems Corporation
- MAGNETIC RESONANCE DIAGNOSTIC ACCESSORY by General Electric Co.
- MRI OE 1.5 SI by Otsuka Electronics
- Vantage Orian 1.5T, MRT-1550, V6.0 with AiCE Reconstruction Processing Unit for MR by Canon Medical Systems Corporation
- VERSION 4.2 APPLICATION SOFTWARE by Resonex, Inc.
- MRT 50A SPINE SURFACE COIL by Toshiba America Mri, Inc.