  "format": "devicerts.device.v1",
  "cite_as": "Optima MR 430S, GE Medical Systems, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/optima-mr-430s-lnh (data as of 2026-11-24).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/optima-mr-430s-lnh",
    "json_url": "https://devicerts.com/devices/optima-mr-430s-lnh.json",
    "markdown_url": "https://devicerts.com/devices/optima-mr-430s-lnh.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-24",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/optima-mr-430s-lnh",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Optima MR 430S",
    "registry_name": "OPTIMA MR 430S",
    "url": "https://devicerts.com/devices/optima-mr-430s-lnh",
    "manufacturer": {
      "name": "GE Medical Systems, LLC",
      "url": "https://devicerts.com/manufacturers/ge-medical-systems-llc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K123068",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "LNH"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K123068",
      "number_label": null,
      "device_udi_di": null,
      "date": "2012-11-13",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0620-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2021-12-23",
      "reason": "The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.",
      "applies": "this device",
      "via": "K123068"
    },
    {
      "number": "Z-1928-2016",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2016-05-09",
      "reason": "The gas venting may not occur properly.  A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.",
      "applies": "this device",
      "via": "K123068"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K123068",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "LNH",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1000",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}