## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K081496 · decision 2008-06-12 · decision SESE · status: Active

## Recalls

- Z-1747-2019 · Class II · ongoing · initiated 2019-02-12: GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Z-1745-2019 · Class II · ongoing · initiated 2019-02-12: GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
- Z-2190-2017 · Class II · terminated · initiated 2017-05-19: It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
- Z-1051-2016 · Class II · terminated · initiated 2016-02-10: GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.
- Z-0981-2014 · Class II · terminated · initiated 2013-11-20: GE Healthcare has recently become aware of a potential safety issue due to scatter overcorrection associated with the List Mode Replay feature of your Discovery 600, 610, 710 and Discovery 690.
- Z-0024-2014 · Class II · terminated · initiated 2013-03-26: GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID
- Z-1188-2013 · Class II · terminated · initiated 2011-07-25 · matched through shared submission K081496, may concern other models: Fastener bolts on the heat exchanger were not properly torqued and part of the heat exchanger can detach from the gantry and hit the inside of the covers and other components of the gantry possibly resulting in the oil conduit breaking, spilling oil within the gantry and outside the system. This issue affects GE Healthcare VCT for Discovery VCT, Discovery RX VCT, and Discovery PET/CT 690, Lightspe

## Identifiers

- 510(k) Number: K081496 [FDA 510(k)]
- FDA product code: KPS [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.1200 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Optima Pet/CT 560, GE Medical Systems, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/optima-pet-ct-560-kps (data as of 2026-11-20).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/optima-pet-ct-560-kps.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/optima-pet-ct-560-kps
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.