Oryx™ AP360
Oryx AP360 is a spinal fixation plate (FDA Class II; FDA product code KWQ, Appliance, Fixation, Spinal Intervertebral Body) made by INNO Holdings, Inc., cleared under FDA 510(k) K192049 on 8 Nov 2019 and listed in FDA GUDID since 31 Jul 2020 (UDI-DI M711AP3600), active as of 23 Oct 2026. One FDA recall is linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | NumberM711AP3600 | Date31 Jul 2020 | StatusListed |
| United States | FDA 510(k)SESE | NumberK192049 | Date8 Nov 2019 | StatusActive |
Official records
- FDA 510(k) database K192049
From the FDA record
- Device description
- Ti 6Al 4V (ELI)
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K192049 decided 8 Nov 2019 Active
- FDA GUDID UDI-DI M711AP3600 listed 31 Jul 2020 Listed
Recalls
-
- Terminated Z-2062-2020
Cervical screw caddies have incorrect text markings for the 4.5mm Fixed and Variable rescue screws. The caddy location labeled as "Fixed" should be labeled as "Variable" and the caddy location labeled as "Variable" should be labeled as "Fixed. If user installs a fixed rescue screw instead of variable, the screw head may not be flush, it could back out, which could interfere with the cover screw.
Timeline
- Cleared FDA 510(k) K192049
- Recall initiated FDA recalls Z-2062-2020
- Listed FDA GUDID M711AP3600
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Oryx™
- Generic name
- Spinal fixation plate, non-bioabsorbable
- Model / Ref / Catalog
- AP360
- Primary DI / UDI-DI
- M711AP3600
- FDA product code
- KWQ
- GMDN term
- Spinal fixation plate, non-bioabsorbable
- Device class
- Class II
- Regulation number
- 888.3060
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Length — 60 Millimeter