Skip to content

Osteonics Ha Generation II Acetabular Component System

by Osteonics Corp.

510(k) K983382

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

4 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK983382 Date11 Dec 1998ClearedStatusActive

Recalls

4 recalls · FDA

    • Terminated Z-1043-2015 FDA Class II initiated 15 Dec 2014

    Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present.

  • 2 recalls, same reason

    • Terminated Z-0073-2008 FDA Unclassified (pre-2012) initiated 30 Aug 2007
    • Terminated Z-0072-2008 FDA Unclassified (pre-2012) initiated 30 Aug 2007

    Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will decrease the gap between shell and liner on the opposing side, resulting in interference.

    • Terminated Z-1261-2007 FDA Unclassified (pre-2012) initiated 13 Mar 2006

    Dimensional anomaly

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MEH
Device Class
Class II
Regulation
21 CFR Part 888.3353
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent

Other devices from Osteonics Corp.