Skip to content

Osteonics Scorpio Posteriorly Stabilized Total Knee System

by Osteonics Corp.

510(k) K962152

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

7 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SN Traditional NumberK962152 Date14 Aug 1996DecidedStatusActive

Recalls

13 recalls · 7 ongoing · FDA

  • 7 recalls, same reason

    • Ongoing Z-0886-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0892-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0887-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0891-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0890-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0888-2024 FDA Class II initiated 1 Dec 2023
    • Ongoing Z-0889-2024 FDA Class II initiated 1 Dec 2023

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    • Terminated Z-1634-2015 FDA Class II initiated 15 Aug 2014

    The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

    • Terminated Z-0469-2010 FDA Unclassified (pre-2012) initiated 2 Oct 2009

    It was determined that a change made in the dimensions of the collar component of the packaging resulted in the inner blister material being overstretched causing a potential hole in the sterile barrier system.

    • Terminated Z-0795-2010 FDA Unclassified (pre-2012) initiated 30 Jul 2007

    Stryker Orthopaedics became aware that a box packaged and labeled as a Scorpio PS Femoral Waffle (No Post) with LFIT, may actually contain a Scorpio Femoral Waffle (With Post) and LFIT.

    • Terminated Z-1814-2008 FDA Unclassified (pre-2012) initiated 8 Nov 2005 on shared 510(k) K962152

    Mislabeled boxes -- Scorpio PS Right Femoral Component No Posts contained a Scorpio PS Left Femoral Component with Posts and vice versa.

    • Terminated Z-2079-2008 FDA Unclassified (pre-2012) initiated 17 Mar 2005

    Scorpio Posterior Stabilized Femoral Component labeled and laser marked as a Left, was in fact a Right component.

    • Terminated Z-1382-2008 FDA Unclassified (pre-2012) initiated 8 Feb 2005

    Mislabeled: The labeling indicates a Scorpio PS Femoral Component, however, the device inside the package is actually a Scorpio CR Femoral Component.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
JWH
Device Class
Class II
Regulation
21 CFR Part 888.3560
Advisory Committee
Orthopedic
FDA Decision
SN

Other devices from Osteonics Corp.

See all devices from Osteonics Corp. →