OXYGEN SENSORS - T1
by Teledyne Analytical Instruments
510(k) K001095
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK001095 | Date30 Jun 2000 | StatusActive |
Recalls
-
- Terminated Z-3001-2011
The recall was initiated because during a test on a batch of sensors on the company shipping dock, Teledyne Instruments discovered electrolyte loss from the sealing cylinder on a few sensors. This test was done pursuant to a complaint from one customer of label discoloration caused because of the electrolyte loss. The risk of a leak form the oxygen sensor while in use is the disruption of therape
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CCL
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.1720
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent