Oxylog 3000plus 5704833
by Dragerwerk AG EUDAMEDGUDID
Withdrawn
Ref 5704833Model 304015FK19Z000
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number040486751304015FK19Z000XW | Date27 May 2026 | StatusWithdrawn |
| United States | GUDID listing GUDID | Number04048675249214 | Date24 Oct 2015 | StatusWithdrawn |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2018-2023
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
-
- Terminated Z-0691-2016
- Terminated Z-0692-2016
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
-
- Terminated Z-1440-2014
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
Specification
- Trade Name
- Oxylog 3000 plus EUDAMEDOxylog 3000plus GUDID
- Generic Name
- Transport electric ventilator
- Model / Reference
- 5704833, 304015FK19Z000
- Primary DI / UDI-DI
- 04048675249214
- Basic UDI-DI
- 040486751304015FK19Z000XW
- Direct Marketing DI
- 04048675249221
- 510(k) Number
- FDA Product Code
- CBK
- EMDN Code
- Z12030104 Portable pulmonary ventilators
- GMDN Term
- Transport electric ventilator
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 868.5895
- Market of Registration
- Germany
- Commercial Distribution
- Not in Commercial Distribution
- MRI Safety
- MR Unsafe
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Device Class | Class IIb | Class II |
Other devices from Dragerwerk AG
Properties
Other properties (19)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Incorporates Medicinal Product: No
- Prescription Use: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No