# PAINPUMP1; PAINPUMP2

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/painpump1-painpump2-frn>
- Manufacturer: [Stryker Corp.](https://devicerts.com/manufacturers/stryker-corp) (FDA registration: United States)
- Device class: Class II (US FDA)
- FDA product code: FRN

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K042405 · decision 2004-10-14 · decision SESE · status: Active

## Recalls

- Z-2591-2010 ·  · terminated · initiated 2010-09-02: All of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a high failure rate associated with E2 errors.
- Z-2592-2010 ·  · terminated · initiated 2010-09-02: All of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a high failure rate associated with E2 errors.
- Z-2593-2010 ·  · terminated · initiated 2010-09-02: All of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a high failure rate associated with E2 errors.
- Z-2594-2010 ·  · terminated · initiated 2010-09-02: All of the failures lead to the patient receiving less medication than intended. Certain lots of the pain pump have a high failure rate associated with a cracked port housing which leads to internal or external leaks. Some lots have a high failure rate associated with E2 errors.
- Z-0096-2011 ·  · terminated · initiated 2010-08-16: The instructions for use (IFU) has been revised for health care professionals. The indication for use has been changed to remove site specific pain management. The firm has revised the warning to not use the pump around joint spaces. A warning has been added regarding the use of the pump with anticoagulants in epidural applications. A special reminder to users to always evaluate the potential
- Z-1736-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1737-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1738-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1739-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1740-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1741-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1742-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1743-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1744-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1745-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1746-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1747-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1748-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1749-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1750-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1751-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1752-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1753-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1754-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1755-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1756-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1757-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1758-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1759-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1760-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1761-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1762-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1763-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1764-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1765-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1766-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1767-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1768-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1769-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1770-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1771-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1772-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1773-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1774-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1775-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1776-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1777-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1778-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1779-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1780-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1781-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1782-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1783-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1784-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1785-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1786-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1787-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-1788-2009 ·  · terminated · initiated 2009-06-16: The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
- Z-0923-2009 ·  · terminated · initiated 2008-11-07: There is a potential for the pump to overinfuse.
- Z-0924-2009 ·  · terminated · initiated 2008-11-07: There is a potential for the pump to overinfuse.
- Z-0925-2009 ·  · terminated · initiated 2008-11-07: There is a potential for the pump to overinfuse.
- Z-0662-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0663-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0664-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0665-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0666-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0667-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0668-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0669-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.
- Z-0670-06 ·  · terminated · initiated 2006-02-03 · matched through shared submission K042405, may concern other models: The catheter tubing included in the kit may break during removal.

## Identifiers

- 510(k) Number: K042405 [FDA 510(k)]
- FDA product code: FRN [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.5725 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: PAINPUMP1; PAINPUMP2, Stryker Corp. deviCERTS medical device catalogue. https://devicerts.com/devices/painpump1-painpump2-frn (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/painpump1-painpump2-frn.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/painpump1-painpump2-frn
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.