Panorama 1.0T

by Philips Medical Systems North America, Inc.

510(k) K041602

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK041602 Date26 Jul 2004ClearedStatusActive

Recalls

7 recalls · 1 ongoing · FDA

    • Ongoing Z-0505-2024 FDA Class I initiated 17 Nov 2023

    Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium gas due to an unforeseen blockage in the system, risks to patients and/or operators that may lead to injury or death include chemical exposure (i.e. helium gas) that may expand to surrounding rooms, contusion caused by debris, and asphyxia.

  • 2 recalls, same reason

    • Terminated Z-0898-2019 FDA Class II initiated 21 Dec 2018
    • Terminated Z-0917-2019 FDA Class II initiated 21 Dec 2018

    There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.

    • Terminated Z-1434-2018 FDA Class II initiated 16 Mar 2018

    Potential risk for helium gas inside the MR examination room during a magnet quench

    • Terminated Z-1458-2011 FDA Unclassified (pre-2012) initiated 5 Feb 2010

    In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of additional X-RAY during treatment.

    • Terminated Z-0794-2011 FDA Unclassified (pre-2012) initiated 12 May 2009

    The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.

    • Terminated Z-0984-2007 FDA Unclassified (pre-2012) initiated 30 Apr 2007

    Potential for gaseous helium to improperly vent, resulting in rupture of the magnet vessel and venting of helium into the MRI site.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LNH
Device Class
Class II
Regulation
21 CFR Part 892.1000
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent