Patello-Femoral Wave (Kahuna) Arthroplasty System
510(k) K181280
Active
510(k)
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181280 | Date14 Jun 2018 | StatusActive |
Recalls
-
- Ongoing Z-1046-2023
- Ongoing Z-1047-2023
- Ongoing Z-1045-2023
The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KRR
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3540
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent