Skip to content

Pathfast Ck-Mb-II, Pathfast Myo-II

by Mitsubishi Kagaku Iatron 510(k)

Active United States

510(k) K081360

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK081360 Date17 Aug 2009ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

1 recall · FDA

    • Terminated Z-3185-2018 FDA Class II initiated 30 Apr 2018

    A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

Specification

510(k)

510(k) Number
FDA Product Code
JHX
Device Class
Class II
Regulation
21 CFR Part 862.1215
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent