PEBA Anchor/Suture Combination, 2.0mm Mini-Tac Anchor, Model 10-1629-01, Modification to Twinfix Ti Quick-T, Twinfix FT PK, Twinfix Ultra Ti, Twinfix Ultra PK, Twinfix Ultra HA Suture Anchors, FOOTPRINT Ultra OK Suture Anchors, BIORAPTOR 2.9 Suture Anchor, BIORAPTOR 2.3 PK Suture Anchor, OSTEORAPTOR Suture Anchor, BIORAPTOR Knotless Suture Anchor, HEALICOIL PK Suture Anchor (formerly Next Generation Fully Threaded Suture Anchor), Bioraptor Curved 2.3 PK Suture Anchors
510(k) K152566
Active
510(k)
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK152566 | Date2 Dec 2015 | StatusActive |
Recalls
-
- Ongoing Z-0581-2025
Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
-
- Terminated Z-1273-2020
Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- MBI
- Models / Variants
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3040
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent