Skip to content

Peba Anchor/Suture Combination

by Orthopaedic Biosystems, Ltd.

510(k) K972326

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK972326 Date6 Feb 1998ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-1273-2020 FDA Class II initiated 30 Dec 2019

    Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MBI
Device Class
Class II
Regulation
21 CFR Part 888.3040
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent

Other devices from Orthopaedic Biosystems, Ltd.