PedFuse 01-80133-30
PedFuse 01-80133-30 is a bone-screw internal spinal fixation system (FDA Class II) made by Spinefrontier, Inc., cleared under FDA 510(k) K123164 on 15 Feb 2013 and listed in FDA GUDID since 24 Sep 2016 (UDI-DI 00190361034087), active as of 23 Oct 2026. 4 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number00190361034087 | Date24 Sep 2016 | StatusListed |
| United States | FDA 510(k)SESE | NumberK123164 | Date15 Feb 2013 | StatusActive |
Official records
- AccessGUDID record 00190361034087
- FDA 510(k) database K123164
From the FDA record
- Device description
- PedFuse Remind LES, Quad, 6.0mm x 30mm
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile; sterilise before use (Moist Heat or Steam Sterilization)
Across registries
- FDA 510(k) K123164 decided 15 Feb 2013 Active
- FDA GUDID UDI-DI 00190361034087 listed 24 Sep 2016 Listed
Recalls
-
- Terminated Z-1479-2015
The PedFuse Pedicle Screw Rod Caliper included with the MISquito Pedicle Screw System may be used incorrectly as the Surgical Technique does not adequately convey the intended use of the design.
-
- Terminated Z-0555-2014
- Terminated Z-0554-2014
- Terminated Z-0553-2014
Screw inserters may not mate properly with hex interface of the screws.
-
- Terminated Z-0544-2014
During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.
-
- Terminated Z-0624-2014
Bone Probes used to create a pathway in pedicle bone were manufactured and marked to incorrect specifications.
Timeline
- Cleared FDA 510(k) K123164
- Recall initiated FDA recalls Z-0624-2014
- Recall initiated FDA recalls Z-0544-2014
- Recall initiated FDA recalls Z-0555-2014
- Recall initiated FDA recalls Z-0554-2014
- Recall initiated FDA recalls Z-0553-2014
- Recall initiated FDA recalls Z-1479-2015
- Listed FDA GUDID 00190361034087
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- PedFuse
- Generic name
- Bone-screw internal spinal fixation system, non-sterile
- Model / reference
- 01-80133-30
- Primary DI / UDI-DI
- 00190361034087
- 510(k) Number
- FDA product code
- MNI, MNH
- GMDN term
- Bone-screw internal spinal fixation system, non-sterile
- Device class
- Class II
- Regulation number
- 888.3070
- GUDID distribution status
- In Commercial Distribution
- Sterilisation method
- Moist Heat or Steam Sterilization
- Size
- Length — 30 MillimeterOuter Diameter — 6 Millimeter
Similar devices
Other devices filed under FDA product code MNI (ORTHOSIS, SPINAL PEDICLE FIXATION).
- Erisma®-LP 13027565 by Clariance, Inc.
- Erisma®-LP 13014535 by Clariance, Inc.
- CD HORIZON SPINAL SYSTEM by MEDTRONIC SOFAMOR DANEK, INC.
- CD HORIZON® Spinal System 55410005550 by MEDTRONIC SOFAMOR DANEK, INC.
- MODIFICATION TO SPINAL CONCEPTS BACFIX SPINAL FIXATION SYSTEM by Spinal Concepts, Inc.
- ROMEO POSTERIOR OSTEOSYNTHESIS SYSTEM by Spineart
- SUMMIT OCT SPINAL SYSTEM MINIPOLYAXIAL SCREWS by Depuyacromed
- MOUNTAINEER 188318318 by MEDOS INTERNATIONAL SARL
- WASTON General Spinal System by Changzhou Waston Medical Appliance Co., Ltd
- MODIFICATION TO SUMMIT OCCIPITO-CERVICO-THORACIC (OCT) SPINAL SYSTEM by Depuyacromed