# Percuvance IPN027877

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/percuvance-pcvsb12-9836a19f>
- Manufacturer: [Teleflex Incorporated](https://devicerts.com/manufacturers/teleflex-incorporated) (FDA registration: United States)
- Name in the registry: PERCUVANCE IPN027877
- Device class: Class II (US FDA)
- FDA product code: GEI
- GMDN: Laparoscopic access cannula seal reducer, single-use (45597)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 14026704582881 · listed 2017-11-16 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K153063 · decision 2016-04-08 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0753-2019 · Class II · terminated · initiated 2018-01-03: There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.
- Z-0754-2019 · Class II · terminated · initiated 2018-01-03: There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.
- Z-0059-2018 · Class II · terminated · initiated 2017-10-18: Teleflex is recalling these products because the jaws of the tool tip may break, causing a fragment of the tool tip to fall off during use.
- Z-3219-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
- Z-3220-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
- Z-3221-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
- Z-3222-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
- Z-3223-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.
- Z-3224-2017 · Class II · terminated · initiated 2017-07-19: For some of the recalled products, the security of the connection between the shaft and the tool tip may be compromised. For certain other recalled lots of the products, the jaws of the tool tips may close abruptly.

## Identity

- Trade name: PERCUVANCE [FDA GUDID (AccessGUDID)]
- Generic name: Laparoscopic access cannula seal reducer, single-use [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: IPN027877 [FDA GUDID (AccessGUDID)]; PCVSB12 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 14026704582881 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K153063 [FDA GUDID (AccessGUDID)]
- FDA product code: GEI [FDA GUDID (AccessGUDID)]
- GMDN term: Laparoscopic access cannula seal reducer, single-use (45597) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 878.4400 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Dimensions & Sizes

- Size: Length — 3.4 Inch [FDA GUDID (AccessGUDID)]; Width — 0.7 Inch [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Percuvance IPN027877, Teleflex Incorporated. deviCERTS medical device catalogue. https://devicerts.com/devices/percuvance-pcvsb12-9836a19f (data as of 2026-11-28).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/percuvance-pcvsb12-9836a19f.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/percuvance-pcvsb12-9836a19f
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.