Perma-Hand C014D is a device classified as Class III in the EU and Class II in the US made by Ethicon Inc., registered in EUDAMED on 15 Sep 2026 (Basic UDI-DI B-USMF000013112000213KC), cleared under FDA 510(k) K946173 on 9 Jan 1995 and listed in FDA GUDID since 1 Sep 2015 (UDI-DI D-USMF000013112000213KC), active as of 24 Oct 2026. 3 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDD 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

5 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

510(k) K946173

FDA submission, from GUDID

Trust

Unverified

Source EUDAMED, GUDID, synced 24 Oct 2026.

Regulatory status

as of registry sync 24 Oct 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDD EUDAMED NumberB-USMF000013112000213KC Basic UDI-DI D-USMF000013112000213KCUDI-DIDate15 Sep 2026RegisteredStatusActive
United StatesGUDID listing GUDID NumberD-USMF000013112000213KC UDI-DI Date1 Sep 2015ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK946173 Date9 Jan 1995DecidedStatusActive

Official records

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI B-USMF000013112000213KC registered 15 Sep 2026 Active
  • FDA 510(k) K946173 decided 9 Jan 1995 Active
  • FDA GUDID UDI-DI D-USMF000013112000213KC listed 1 Sep 2015 Listed

Recalls

10 recalls · 5 ongoing · FDA

    • Ongoing Z-0284-2027 FDA Not yet classified initiated 23 Sep 2026 on shared 510(k) K946173

    Nylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.

    • Ongoing Z-0153-2027 FDA Class II initiated 8 Sep 2026 on shared 510(k) K946173

    Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.

    • Ongoing Z-1516-2025 FDA Class II initiated 7 Mar 2025 on shared 510(k) K946173

    Identified curing issues with the silicone during the needles manufacturing process.

    • Ongoing Z-1097-2025 FDA Class II initiated 20 Dec 2024 on shared 510(k) K946173

    Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.

    • Ongoing Z-2486-2020 FDA Class II initiated 10 Jun 2020

    Lots of size 8-0 Blue twisted Silk Suture may not meet the stringent full shelf life of 5 years

  • 2 recalls, same reason

    • Terminated Z-1841-2019 FDA Class II initiated 19 Apr 2019 on shared 510(k) K946173
    • Terminated Z-1842-2019 FDA Class II initiated 19 Apr 2019 on shared 510(k) K946173

    Possibility that a suture raw material containing high endotoxin levels was used in the manufacturing of this lot

    • Terminated Z-0747-2019 FDA Class II initiated 18 Dec 2018

    The sutures inside the packaging are not the same size or type as indicated on the label.

    • Terminated Z-1839-2015 FDA Class II initiated 8 May 2015 on shared 510(k) K946173

    Ethicon Fast Absorbing Surgical Gut (Plain) Suture is being recalled because the labeling on the box and foil are incomplete.

    • Terminated Z-0492-2015 FDA Class II initiated 17 Oct 2014

    Ethicon is unable to ensure that PERMA-HAND silk suture (siliconized black braid) meets the expected sterility assurance level for sterile products.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

13 dated events

  1. Listed FDA GUDID D-USMF000013112000213KC
  2. Recall initiated FDA recalls Z-0747-2019
  3. Recall initiated FDA recalls Z-1841-2019
  4. Recall initiated FDA recalls Z-1842-2019
  5. Recall initiated FDA recalls Z-2486-2020
  6. Recall initiated FDA recalls Z-1097-2025
  7. Recall initiated FDA recalls Z-1516-2025
  8. Recall initiated FDA recalls Z-0153-2027
  9. Registered EUDAMED B-USMF000013112000213KC
  10. Recall initiated FDA recalls Z-0284-2027

3 earlier events are in the sections above.

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDD Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
Perma-Hand
Generic name
Silk suture
Model / Ref / Catalog
C014D
Primary DI / UDI-DI
D-USMF000013112000213KC EUDAMED
10705031023451 GUDID
Basic UDI-DI
B-USMF000013112000213KC
510(k) Number
FDA product code
GAP
EMDN code
H010202020301 Silk with needle
GMDN term
Silk suture
Device class
Class III EUDAMED
Class II GUDID
Regulation number
878.5030
EU registration type
Legacy device (under a directive)
EU market of registration
Belgium
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 2 fields differ

FieldEUDAMEDGUDID
Primary DI / UDI-DIEUDAMED
D-USMF000013112000213KC
GUDID
10705031023451
Device classEUDAMED
Class III
GUDID
Class II

The registries describe this device under different primary UDI-DIs (EUDAMED lists D-USMF000013112000213KC; FDA GUDID lists 10705031023451), so their flags are not compared. Device class is stated on each registry's own scale: Class III in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under SILK WITH NEEDLE.

Other devices from Ethicon Inc.

See all devices from Ethicon Inc. →

Properties

  • Supplied sterile
  • Single use
  • Implantable
  • Prescription use
  • Lot / batch number
  • Expiration date
Other properties (19, all answered No)