Perseus A500
by Draeger Medical GmbH 510(k)
Active
510(k) K133886
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK133886 | Date20 Jun 2014 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1512-2024
Unexpected shutdown while operating on battery power.
-
- Terminated Z-2254-2018
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
-
- Terminated Z-0360-2016
There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The workstation has a fresh gas delivery failure and a ventilator failure with respective alarms. 2. The workstation shuts down completely with an audible alarm.
Specification
- 510(k) Number
- FDA Product Code
- BSZ
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5160
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent