Pharos Excimer Laser System, Model Ex-308
510(k) K062963
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK062963 | Date3 Apr 2007 | StatusActive |
Recalls
-
- Terminated Z-1631-2020
At doses of less than 100 mJ, in custom and auto repeat modes, a software error in the excimer laser phototherapy system causes the device to fail to cease firing as long as the foot pedal or handpiece switch is depressed. This malfunction can cause overtreatment, which could result in strong erythema, and possibly cause blisters.
-
- Terminated Z-2123-2014
RA Medical Systems is recalling the PHAROS EX-308 Excimer Laser because there is a possibility that the laser may calibrate with the iris closed.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- GEX
- FDA applicant
- Ra Medical Systems,Inc
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4810
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent